SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
Single / Continuous / Restart
This setting determines whether the FieldFox measures continuously or only
once each time the
Single or to conserve battery power or to allow you to save or analyze a
specific trace.
How to set Single or Continuous
1. Press Sweep 3.
2. Then choose one of the following:
Points
Points is the number of measurements that are displayed along the X-axis. The
higher number of data points, the better the ability to resolve closely spaced
signals and the slower the sweep speed.
How to set number of points
1. Press Sweep 3 >
2. Then choose from:
Trace Display States (SA Mode)
— In SA Mode you can display up to four of the following types of trace states.
Keysight N9938-90006 User's Guide
Single
—
Continuous
—
ON
Causes the SA to make continuous sweeps. This is the
typical setting for making measurements. Each Run / Hold
+/- press causes the SA to toggle between continuous
sweep and Hold.
—
OFF
Causes the SA to Hold. Each Run / Hold +/- or
press causes the SA to make ONE sweep, then Hold. The
Hold annotation changes to an arrow --> while the sweep
occurs.
—
Single
Automatically sets Continuous OFF and causes the SA to
make ONE sweep, then Hold.
—
Restart
Resets the Average counter to 1. The Average counter also
is used with MaxHold and MinHold Trace States. Learn more in
"Average Count" on page
Points [current setting]
—
101 | 201 | 401 | 601 | 801 | 1001
—
# Points
to enter a trace points value manually (Maximum = 20001).
All SA settings are applied to all displayed traces.
(IQA and SA Mode)" on page
or Run / Hold +/- button is pressed. Use Hold /
203.
730.
Single
See also
"Marker Trace
201