Digital Analyzer > Meas Config > Trigger menu description
Menu
Description
Press the Trigger Source softkey to select the trigger source. The default trigger in is Free Run, where the analyzer triggers
immediately after the previous data is acquired. If External trigger in is set, the analyzer waits for a trigger pulse on the
Trigger In connector at the rear panel before acquiring the measurement data.
Trigger Source
– Free Run
– External
– Channel
This is only available when the trigger source is External or Channel.
Press the Trigger Edge softkey to select the trigger edge type.
Trigger Edge
– Rising
– Falling
This is only available when the trigger source is Channel.
Press the Interface softkey to select the trigger interface.
Interface
– Analog
– Digital
– Bluetooth
This is only available when the trigger source is Channel.
Channel
Press the Channel softkey to set the channel number.
This is only available when the trigger source is Channel.
When the Trigger Interface is Analog, the Trigger Level is in Vpeak.
Trigger Level
When the Trigger Interface is Digital or Bluetooth, the Trigger Level is in Hex.
Press the Trigger Level softkey to set the trigger level.