Sweep; Table 8-5 Hp8903B > Sweep Menu Description - Keysight U8903B User Manual

Performance audio
Hide thumbs Also See for U8903B:
Table of Contents

Advertisement

Sweep

In HP8903B mode, the source frequency is logarithmically swept. The number of frequency
points in a sweep is determined by the sweep width (ratio of the stop and start frequencies)
and the selected sweep resolution. The maximum number of points allowable in a sweep is
255. The frequency points in a sweep can be computed by using the following formulas:
Frequency
Where n is the frequency point number (0 is the start frequency) and k is the number of
points per decade. Refer to the HP8903B special function codes, 17.0 to 17.9 as listed in
Table 8-6. The frequency point formula for reverse sweep is as follows:
Frequency
Select Sweep in the HP8903B menu page to configure the HP8903B sweep settings.
Figure 8-4
Table 8-5
HP8903B > Sweep menu description
Menu
Description
Freq. Start
Press the Freq. Start softkey to set the HP8903B sweep frequency start value.
Freq. Stop
Press the Freq. Stop softkey to set the HP8903B sweep frequency stop value.
Sweep Status
Press the Sweep Status softkey to start or abort sweeping in the HP8903B mode.
Keysight U8903B User's Guide
n
⎛ ⎞
- -
⎝ ⎠
k
=
Start frequency 10
×
- - - -
k
=
Start frequency 10
×
HP8903B > Sweep menu page
n
HP8903B
8
413