Keysight U8903B User Manual page 502

Performance audio
Hide thumbs Also See for U8903B:
Table of Contents

Advertisement

A
Appendixes
Table A-14
Test sequence menu tree description (continued)
Tab
Level 1
Measurement
Properties
502
Level 2
Level 3
Sub-Steps:
Properties
Prompt
Enable
Add Sub-Step
Delete Sub-Step
Edit
Sub-Steps:
Send SCPI
Settings
Properties
– Cancel Seq.
Failure Handling
– Allow Retry
– Continue Seq.
Description
Set the prompt sub-step name.
Enable or disable the send SCPI sub-step.
Add sub-step to the list.
– Delay
– Prompt
– Sent SCPI
Delete the selected sub-step.
Move or copy the selected sub-step, or paste a
copied sub-step after the selected sub-step.
– GPIB Address
Select the desired GPIB address.
– SCPI Commands
– Edit
Set the SCPI commands at the lower half of
the main display.
– Import
Load the SCPI commands from a file. Refer
to "Append ix 13: Recall Menu Tree" on
page
494
for the recall menu tree.
– Progress Msg
Set the optional text message that will be
displayed on a dialog box for the time length
set in Delay.
– Delay
Set the delay time length after the SCPI
command(s) is executed.
Set the prompt sub-step name.
Select the failure handling type for the selected
measurement.
Keysight U8903B User's Guide

Advertisement

Table of Contents
loading

Table of Contents