To measure the voltage of the sine waveform, perform the following steps.
1 Press
2 Press Analysis Mode, and select Standard as the digital analyzer analysis mode.
3 Press Input Config > Connector, and select Unbalanced as the digital analyzer input type.
4 Press Functions > Function No., and select 1.
5 Press Functions > Function 1, and select Frequency as the first measurement function.
6 Press Functions> Function No., and select 2.
7 Press Functions > Function 2, and select AC Vol tage as the second measurement function.
8 Press
You should now obtain an AC voltage reading of 1 FFS for the generated sine waveform,
within the tolerance as stated in "Characteristics and Specifications" on page 417. The
U8903B display should look as follows.
Figure A-10
572
on the FUNCTION panel to switch to audio analyzer mode, and press
on the FUNCTION panel to switch to digital interface.
to start the signal measurement on the digital analyzer channel 1.
Digital analyzer voltage measurement
Keysight U8903B User's Guide