7
Test Sequence Application

DC level sweep

DC level sweep measurement sweeps a DC signal across a range of values in a series of
points and the DUT output is acquired by the analyzer. The results are displayed on an X-Y
graph, with the generator DC level on the X-axis and the measured result on the Y-axis. If
AC coupling is selected in the input configuration settings, it will temporary be set to DC
coupling while the DC level sweep measurement is performed.
Refer to "Measurement Results" on page
menu page. Use the arrow keys to navigate through the respective tabs and to display the
menu page.
This measurement requires a closed-loop configuration in which both the generator and
NOTE
the analyzer are used in the test setup. If None is selected for the output configuration
channel, this measurement is unavailable.
Signal generation
The signal generation is disabled if None is selected for the output configuration channel.
NOTE
The DC level sweep signal generation settings menu page is displayed as shown in
Figure 7-51.
Figure 7-51
376
DC Level Sweep > Signal Generation settings menu page
399
for more information on the respective result
Keysight U8903B User's Guide