Audio Analyzer Measurement Functions
Table 4-18
Digital Analyzer > Functions > Positive Peak menu description (continued)
Menu
Description
Press the Format softkey to select the format of the returned measurement reading.
– Off
Selecting Off will return the measurement reading without any formatting (raw value).
– Logarithmic
Selecting Logarithmic will return the measurement result by using the following formula:
Format
20
The result is returned in unit dBr.
– Linear
Selecting Linear will return the measurement result by dividing the measured level by the reference level. The result is
returned in unit x.
Set resul t as ref. from
Press the Set resul t as ref. from softkey to store the measurement result from the selected channel as the reference level.
Ref. Level
Press the Ref. Level softkey to set the reference level value. There is only one reference level for each channel.
Vol t/FS
Press the Vol t/FS softkey to set the volts/FS reference value.
Cal SPL
Press the Cal SPL softkey to set the calibrator level value. This setting is only available when the unit is set to dBSPL.
222
Measured rms voltage
- - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - -
log
10
Reference level
Keysight U8903B User's Guide