Illegal Parameter Value; Gated Lo Is Not Available While Sweep Type Is Set To Fft - Keysight X-Series Manual

Signal analyzers, instrument messages
Hide thumbs Also See for X-Series:
Table of Contents

Advertisement

Instrument Messages
Event Messages
Err#
Message
-224
Illegal parameter
value; Gated LO is not
available while Sweep
Type is set to FFT
-224
Illegal parameter
value; Gated Video is
not available while
Sweep Type is set to
FFT
-224
Illegal parameter
value; Index out of
range
-224
Illegal parameter
value; Invalid list
length
-224
Illegal parameter
value; Measurement not
available
-224
Illegal parameter
value; new Name
already exists
-224
Illegal parameter
value; QPD + EMI
Average + RMS Average
is not allowed
-224
Illegal parameter
value; Res BW value
not allowed with
current Span
-224
Illegal parameter
value;Cannot set
waveform to continue
previous waveform when
previous step is set
to CW or Off
-224
Illegal parameter
value;Channel number
is invalid for current
band/link combination
-224
Illegal parameter
value;This instrument
is always DC coupled
Instrument Messages
Verbose/Correction Information
The FFT sweep type moves the LO frequency in steps. So the gated LO
function is not available if you have selected FFT sweep.
The gated video function is not available if you have selected the FFT
sweep type.
When querying the LXI Event Log or the Servo Log, an index may be
used to look at a specific entry. This error occurs if the index provided
does not point to a valid entry.
You are trying to set some list measurement settings, but the multiple
lists that you sent were not all the same length. The number of settings
must be consistent from list to list.
You tried to turn on a measurement that is not available in the current
mode.
An attempt was made to specify a new Name, but the name already
exists; a unique name must be specified.
User is not allowed to turn on all 3 EMI detectors together. You must
turn off one of the EMI Detectors before you turn this on.
With Option B5X, with Spans greater than 255 MHz the Res BW
desired is not available.
An attempt was made to set the step duration for a step in a sequence
to "Continue Previous", but the previous step was outputting a CW
tone or is Off, so there is no waveform to continue.
An attempt was made to enter a channel number for a step within the
List Sequencer, but the channel number is out of the valid range for
the Band and Radio Band Link direction combination defined within
the step.
You can't set AC coupling in this instrument