Mmemory:store:state[:Register]; Output:dc:isolation:level:auto - Keysight E4980A User Manual

Precision lcr meter
Hide thumbs Also See for E4980A:
Table of Contents

Advertisement

SCPI Command Reference
E4980A/AL commands
Description
Loads the state from the memories. Numbers 0 to 9 are located in the internal
memory, while Nos. 10 to 19 are situated in the external USB memory. (No
Query)
Parameter
Range
Resolution
Equivalent key
[Save/Recall] - No. - RECALL

:MMEMory:STORe:STATe[:REGister]

Syntax
:MMEMory:STORe:STATe[:REGister] <numeric>
Description
Stores the state to the memories. Numbers 0 to 9 are located in the internal
memory, while Nos. 10 to 19 are situated in the external USB memory. Number
10 is automatically recalled. (No Query)
Parameter
Range
Resolution
Equivalent key
[Save/Recall] - No. - SAVE

:OUTPut:DC:ISOLation:LEVel:AUTO

Syntax
:OUTPut:DC:ISOLation:LEVel:AUTO {ON|OFF|1|0}
:OUTPut:DC:ISOLation:LEVel:AUTO?
Description
Enables the auto range function for DCI.
Parameter
ON or 1 (Preset value)
OFF or 0
Equivalent key
[Meas Setup] - MEAS SETUP - DCI RNG - AUTO
Keysight E4980A/AL Precision LCR Meter
<Numeric>
0 to 19
1
<Numeric>
0 to 19
1
Description
Enables the auto range function for DCI
Disables the auto range function for DCI
361