Meas - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

MEAS

MEAS
Syntax
MEAS {IMPH|IRIM|LSR|LSQ|CSR|CSQ|CSD|AMPH|ARIM|LPG|LPQ|CPG|CPQ|CPD|COMP
|IMLS|IMCS|IMLP|IMCP|IMRS|IMQ|IMD|LPR|CPR}
MEAS?
Description
Selects impedance measurement parameters as a pair for trace A and trace B. Selecting
COMP specifies the vector measurement; selecting others, the scalar measurement. There
are functional differences between the vector measurement and the scalar measurement
including available display formats and trace operations.
Parameters
For details on each measurement parameter, refer to "Setting Measurement Parameters" on
page 36.
Query response
{IMPH|IRIM|LSR|LSQ|CSR|CSQ|CSD|AMPH|ARIM|LPG|LPQ|CPG|CPQ|CPD|COMP|I
MLS|IMCS|IMLP|IMCP|IMRS|IMQ|IMD|LPR|CPR}<newline><^END>
348
IMPH (initial value) |Z|
IRIM
R
LSR
Ls
LSQ
Ls
CSR
Cs
CSQ
Cs
CSD
Cs
AMPH
|Y|
ARIM
G
LPG
Lp
LPQ
Lp
CPG
Cp
CPQ
Cp
CPD
Cp
COMP
Z
IMLS
|Z|
IMCS
|Z|
IMLP
|Z|
IMCP
|Z|
IMRS
|Z|
IMQ
|Z|
IMD
|Z|
LPR
Lp
CPR
Cp
Set measurement parameters
Trace A
θ
X
Rs
Q
Rs
Q
D
θ
B
G
Q
G
Q
D
Y
Ls
Cs
Lp
Cp
Rs
Q
D
Rp
Rp
Trace B
Chapter 16

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents