GPIB Command Reference

DCOMLOAD{R|L}

DCOMLOAD{R|L}
Syntax
DCOMLOADR <numeric>[OHM]
DCOMLOADL <numeric>[H]
DCOMLOAD{R|L}?
Description
For fixture compensation data measurement ("COM{A|B|C}" command on page 281), sets
the resistance value ("DCOMLOADR" command) and the inductance value
("DCOMLOADL" command) of the LOAD standard used to measure LOAD
compensation data.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Cal]
key
292
Description
Resistance value or inductance value
Range
For resistance value: -100E9 to 100E9
For inductance value: -1E6 to 1E6
Initial value
For resistance value: 50
For inductance value: 0
For resistance value: Ω (ohm)
Unit
For inductance value: H (henry)
Resolution
1E-12
-
FIXTURE COMPEN
<numeric>
-
-
DEFINE VALUE
{LOAD RESIST (R)
|
LOAD INDUCT (L)}