LIMLSTOP

Syntax
LIMLSTOP <numeric>[DEG|RAD|OHM|SIE|H|F|PCT]
LIMLSTOP?
Description
In the creation/edit of the limit line table, sets the lower limit of the stop point of each
segment (specified with the "LIMSTOP" command on page 336). Use this command,
together with the "LIMUSTOP" command on page 338, when specifying the limits using
the upper limit and the lower limit.
If segment creation/edit has not been started, executing this command automatically starts
the edit of the edit target segment and then sets the lower limit of the stop point.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit, starts the edit of the edit target segment, and then
sets the lower limit of the stop point.
Parameters
Description
Range
Initial value
Unit
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[System]
key
Chapter 16
Lower limit of the segment stop point
-1E9 to 1E9
When creating segment 1: 0
When creating an additional segment: Set value of the previous
segment
Varies depending on the measurement parameter (refer to the
explanation of unit in "SCAL" on page 423).
-
-
LIMIT TEST
EDIT LIMIT LINE
GPIB Command Reference
<numeric>
-
-
-
EDIT
more 1/3