CALIBRATION ABORTED
One of the following occurred.
In the middle of or after the completion of the setup for user calibration or fixture
compensation (measurement of required calibration data and calculation and storage of
error coefficients), the setting of calibration points (FIXED or USER) was changed.
The current setup and error coefficients previously stored are now invalid.
When the setting of calibration points was USER, in the middle of or after the
completion of the setup for user calibration or fixture compensation (measurement of
required calibration data and calculation and storage of error coefficients), the sweep
condition (sweep range, sweep parameter, number of sweep points, sweep type) was
changed. The current setup and error coefficients previously stored are now invalid.
In the middle of the setup for user calibration or four-terminal pair extension
(measurement of required calibration data and calculation and storage of error
coefficients), the setup was canceled (the cancel key, CALQUI command, or
ECALQUI command was executed). The setup is invalid. However, error coefficients
previously stored are available.
If necessary, perform the setup for user calibration, fixture compensation, or four-terminal
pair extension.
17
BACKUP DATA LOST
The contents of the battery-backed memory (SRAM), storing user calibration data, fixture
compensation data, GPIB address, IP address, and so on, were lost, causing an error (check
sum error). Its cause is running out of the battery for backup. To enable the battery backup
for the memory again, turn on the power and keep it approximately 10 minutes to charge
the battery completely.
19
UNEXPECTED DATA DETECTED
During the measurement of calibration data (OPEN, SHORT, LOAD) for user calibration,
fixture compensation, and four-terminal pair extension setting, abnormal data was detected
and the acquisition of the calibration data was aborted. Check to see if there is no problem
in standards or setup you use.
26
PRINTER: not on, not connected, out of paper
The printer does not respond to the control from the 4294A. Check the power to the printer,
online status, paper, and so on. Or, the connected printer may not be supported. For
information on supported printers, refer to the "Options and accessories" chapter.
34
MEMORY TRACE NOT ACTIVE
Though no data was stored in the memory trace, you attempted to execute a command
(DISP MEMO, DMNM, and so on) that used the memory trace. The command was invalid.
First, use the DATMEM command and store data in the memory trace.
35
CAN'T CALCULATE EQUIVALENT PARAMETERS
Because measurement data does not match with the selected equivalent circuit, equivalent
circuit parameters cannot be calculated. Acquire data again or select another equivalent
circuit.
36
MUST BE MORE THAN 2 POINTS FOR ANALYSIS
Though the calculation of equivalent circuit parameters (CALCULATE PARAMETERS
Appendix D
Error messages
Error number: 13