Statistics Analysis - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

Processing Measurement Results
Analyzing Measurement Results
90
100
110
120
130
140
150
160
170
180
190
200
210
220
230
240
250
260
270
280
290
300
310
320
330
340
350
360
370
380
390
400
410
420
430

Statistics analysis

The Agilent 4294A allows you to analyze statistics (average, standard deviation, difference
between maximum and minimum). Use the following command to turn this function ON or
OFF.
NOTE
The statistics analysis function is available when the marker function is ON.
The applicable range for statistics analysis is the same as the range searched by the marker.
You can execute the above command to display analysis results. Use the following
command to read analysis results.
102
Cent=6.0E+7
Span=2.0E+4
Model$="E"
!
OUTPUT @Hp4294a;"PRES"
OUTPUT @Hp4294a;"MEAS IMPH"
OUTPUT @Hp4294a;"CENT ";Cent
OUTPUT @Hp4294a;"SPAN ";Span
OUTPUT @Hp4294a;"SING"
OUTPUT @Hp4294a;"*OPC?"
ENTER @Hp4294a;Buff$
!
OUTPUT @Hp4294a;"TRAC A"
OUTPUT @Hp4294a;"AUTO"
OUTPUT @Hp4294a;"TRAC B"
OUTPUT @Hp4294a;"AUTO"
OUTPUT @Hp4294a;"EQUC CIR"&Model$
OUTPUT @Hp4294a;"CALECPARA"
OUTPUT @Hp4294a;"DEFECR1?"
ENTER @Hp4294a;R1
OUTPUT @Hp4294a;"DEFECC1?"
ENTER @Hp4294a;C1
OUTPUT @Hp4294a;"DEFECL1?"
ENTER @Hp4294a;L1
OUTPUT @Hp4294a;"DEFECC0?"
ENTER @Hp4294a;C0
!
PRINT " --- Result ---"
PRINT "R1 :";R1;"[ohm]"
PRINT "C1 :";C1;"[F]"
PRINT "L1 :";L1;"[H]"
PRINT "C0 :";C0;"[F]"
!
OUTPUT @Hp4294a;"SIMFCHAR"
END
"MEASTAT" on page 349
"OUTPMSTA?" on page 388
Chapter 7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents