Scal - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

SCAL

Syntax
SCAL <numeric>[OHM|DEG|RAD|SIE|H|F|PCT]
SCAL?
Description
Sets the value of 1 grid tick in the Y axis (vertical axis) direction in the linear Y axis format
or the value of 1 grid tick in the X axis (horizontal axis) and Y axis directions in the
complex plane format.
Parameters
Description
Range
Initial value
Unit
Resolution
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Scale Ref]
key
Chapter 16
Value of 1 grid tick
1E-15 to 100E6
For the linear Y axis format:
Varies depending on the measurement parameter as follows:
For |Z|, R, X, Rs, Rp: 100E3
For θ: 36
For |Y|, G, B, D: 100E-3
For Cs, Cp: 100E-6
For Ls, Lp: 1
For Q: 100
For the complex plane format:
50
Varies depending on the measurement parameter as follows:
For |Z|, Z, R, X, Rs, Rp: Ω(ohm)
For θ: °(degree)
For |Y|, Y, G, B: S (siemens)
For Cs, Cp: F (farad)
For Ls, Lp: H (henry)
For Q, D: No unit
*2
1E-15
*1.Can be changed to radian using the
*2.This is the minimum value (when the set value is small). The resolution
becomes larger as the set value becomes larger.
-
SCALE/DIV
GPIB Command Reference
<numeric>
*1
"PHAU" command on page 398
SCAL
.
423