CALECPARA

Syntax
CALECPARA
Description
Executes the equivalent circuit analysis depending on the equivalent circuit model
specified with the "EQUC" command on page 311, and displays the analysis result
(equivalent circuit parameters). The analysis range is the same as that of the marker search
function. (No query)
Corresponding
[Display]
key

CALP

Syntax
CALP {FIXED|USER}
CALP?
Description
Specifies which measurement points are used for user calibration data and fixture
compensation data: (fixed) measurement points provided by the 4294A or user created
measurement points (measurement points in the sweep setting at the execution of
calibration/compensation).
Parameters
FIXED
(Initial value)
USER
Query response
{FIXED|USER}<newline><^END>
Corresponding
[Cal]
key

CALQUI

Syntax
CALQUI
Description
Aborts the measurement of user calibration data. (No query)
Corresponding
[CAL]
key

CALS

Syntax
CALS
Description
Lists the set values of the user calibration kit (values entered with the
"DCALOPEN{G|C}"(288page), "DCALSHOR{R|L}"(289page), and
"DCALLOAD{R|L}"(287page) commands). (No query)
Corresponding
[Copy]
key
Chapter 16
-
-
more 1/2
EQUIV CKT
Uses the (fixed) measurement points provided by the 4294A.
Uses the user created measurement points.
-
COMP POINT [ ]
-
USER CAL
-
EXCUTE CAL
CAL KIT VALUE 
-
SELECT CONTENTS
-