SMKRP{1-7}

Syntax
SMKRP{1|2|3|4|5|6|7} <numeric>
SMKRP{1|2|3|4|5|6|7}?
Description
For the active trace (set with the "TRAC" command on page 457), moves a sub marker
(marker number 1-7) to the specified measurement point. In the coupled marker mode (set
with the "MKRCOUP" command on page 352), regardless of the active trace setting,
moves the sub marker on both the A and B traces. If this command is executed as Query, it
reads out the measurement point number at the sub marker position. If the sub marker is
off, executing this command causes an error and the command is ignored.
Parameters
Description
Range
Initial value
Resolution
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
No front panel key is available to execute this function.
key
Chapter 16
<numeric>
Number of the measurement point where you want to move the sub
marker (the number assigned in order from the left edge assuming
that the number of the leftmost measurement point is 1)
1 to the number of points
1
1
GPIB Command Reference
SMKRP{1-7}