Searmax - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

SEARMAX

Syntax
SEARMAX <numeric>[HZ|MHZ|V|A]
SEARMAX?
Description
For the marker search function for the active trace (set with the "TRAC" command on page
457), if the partial search is enabled (specified to ON with the "PARS" command on page
395), sets the partial search range upper border value.
Parameters
OSC level (voltage)
OSC level (current)
dc bias level (voltage)
dc bias level (current)
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Search]
key
Chapter 16
Description
Range
When the sweep
Initial value 110E6
parameter is
Unit
frequency
Resolution 1E-3
Range
When the sweep
Initial value 1
parameter is
Unit
Resolution 1E-3
Range
When the sweep
Initial value 20E-3
parameter is
Unit
Resolution 1E-6
Range
When the sweep
Initial value 40
parameter is
Unit
Resolution 1E-3
Range
When the sweep
Initial value 100E-3
parameter is
Unit
Resolution 20E-6
-
SEARCH RANGE MENU
GPIB Command Reference
<numeric>
Upper border value of the partial search range
40 to 110E6
Hz
5E-3 to 1
V (volt)
200E-6 to 20E-3
A (ampere)
-40 to 40
V (volt)
-100E-3 to 100E-3
A (ampere)
-
MAXIMUM
SEARMAX
429

Hide quick links:

Advertisement

Table of Contents
loading