Agilent Technologies 4294A Programming Manual page 67

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

1080
IF UPC$(Inp_char$)="Y" THEN
1090
ON INTR Scode GOTO Meas_end
1100
ENABLE INTR Scode;2
1110
SELECT Standard$
1120
CASE "Open"
1130
OUTPUT @Hp4294a;"COMA"
1140
CASE "Short"
1150
OUTPUT @Hp4294a;"COMB"
1160
CASE "Load"
1170
OUTPUT @Hp4294a;"COMC"
1180
END SELECT
1190
PRINT "Now measuring..."
1200 Meas_wait: GOTO Meas_wait
1210 Meas_end: !
1220
PRINT Standard$&" Data Measurement Complete"
1230
RETURN 0
1240
ELSE
1250
PRINT "Program Interruption"
1260
RETURN -1
1270
END IF
1280 FNEND
1290
!
1300
! Data Input Function
1310
!
1320 SUB Inp_data(Mes$,Inp_val)
1330
DIM Inp_char$[30]
1340
ON ERROR GOTO Inp_start
1350 Inp_start:!
1360
PRINT "Input "&Mes$
1370
INPUT "Value?",Inp_char$
1380
Inp_val=VAL(UPC$(Inp_char$))
1390
PRINT "Input value: ";Inp_val
1400
INPUT "OK? [Y/N]",Inp_char$
1410
IF UPC$(Inp_char$)<>"Y" THEN Inp_start
1420
OFF ERROR
1430 SUBEND
Chapter 4
Preparing For Accurate Measurement
Fixture Compensation
67

Hide quick links:

Advertisement

Table of Contents
loading