Calst; Cbri - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

CALST

CALST
Syntax
CALST {ON|OFF|1|0}
CALST?
Description
Toggles on and off the user calibration function. If the user calibration data is not stored
(executed with the "CALDON" command on page 274), executing this command to turn
on the calibration causes an error and the command is ignored.
Parameters
Query response
{1|0}<newline><^END>
Corresponding
[Cal]
key

CBRI

Syntax
CBRI <numeric>[PCT]
CBRI?
Description
Sets the brightness of the display color of the item selected with the "COLO" command on
page 280.
Parameters
Query response
{numeric}<newline><^END>
Corresponding
[Display]
key
BRIGHTNESS
276
ON or 1
OFF or 0 (initial value)
-
-
USER CAL
CORRECTION on OFF
Description
Percentage of brightness (the brightest state: 100 %)
Range
0 to 100
Initial value
Varies depending on the display item selected with the "COLO"
command on page 280.
Unit
% (percentage)
Resolution
1
-
more 1/2
-
MODIFY COLOR
Description
Turns on the user calibration function.
Turns off the user calibration function.
<numeric>
-
MODIFY ITEM COLOR
-
MODIFY COLOR
-
Chapter 16