Limstop - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

LIMSTOP

LIMSTOP
Syntax
LIMSTOP <numeric>[HZ|MHZ|V|A]
LIMSTOP?
Description
In the creation/edit of the limit line table, sets the sweep parameter value of each segment
stop point.
If segment creation/edit has not been started, executing this command automatically starts
the edit of the edit target segment and then sets the sweep parameter value of the stop point.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit, starts the edit of the edit target segment, and then
sets the sweep parameter value of the stop point.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[System]
key
336
Description Sweep parameter value of the segment stop point
Range
same as "STOP" command on page 446
Initial value When creating segment 1: same as "STOP" command
When creating an additional segment: Set value of the previous segment
Unit
same as "STOP" command
Resolution
same as "STOP" command
-
LIMIT TEST
-
<numeric>
EDIT LIMIT LINE
-
EDIT
-
more 1/3
-
STOP
Chapter 16

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents