Error messages
Error number: 114
Though you attempted to display the list of files and directories on a floppy disk on the
softkey label by executing the recall ([Recall]) key, file resave (RE-SAVE FILE) key, file
deletion (PURGE FILE) key, directory change (CHANGE DIRECTORY) key, or file copy
(COPY FILE) key, no file or directory corresponding to the execution of the key was on
the floppy disk.
114
CAN'T SAVE GRAPHICS WHEN COPY IN PROGRESS
You commanded to save the graphics screen during the execution of copying (outputting to
a printer), but the graphics screen cannot be saved during output to a printer.
116
NO STATE/DATA FILES ON MEMORY
Though you attempted to display the list of files and directories on a volatile memory disk
or nonvolatile memory disk (flash memory) on the softkey label by executing the recall
([Recall]) key, file resave (RE-SAVE FILE) key, file deletion (PURGE FILE) key,
directory change (CHANGE DIRECTORY) key, file copy (COPY FILE) key, and so on,
no file or directory corresponding to the execution of the key was on the volatile memory
disk or nonvolatile memory disk (flash memory).
118
DEVICE TYPE IS NOT DOS
The execution of write access to the mass storage could not be executed because the format
type of the mass storage was not DOS.
124
LIST TABLE EMPTY OR INSUFFICIENT TABLE
When the list sweep table was empty or insufficient, you attempted to select LIST as the
sweep type (command: SWPT LIST) and, as a result, the command was ignored. Before
selecting the list sweep as the sweep type, set the list sweep table correctly.
130
CAN'T CHANGE ON ZERO SPAN
During the execution of the list sweep when the sweep span of each segment was zero, you
attempted to change the list sweep span (integrated total span of individual segments
specified as the horizontal axis of the graph for the list sweep) to the single span (the way
to specify a linear frequency axis for the horizontal axis of the graph whose left edge and
right edge are the minimum frequency and the maximum frequency within all segments)
using the LIST SPAN [ ] key or the LISPAN SINGLE command and, as a result, the
command was ignored. When executing the list sweep while the sweep span of each
segment being zero, the list sweep span is automatically set to the segment span and cannot
be changed.
131
FREQUENCY SWEEP ONLY
When a sweep parameter other than the frequency was specified, you attempted to execute
an equivalent circuit analysis command (CALECPARA and so on). You can execute the
equivalent circuit analysis only when the sweep parameter is the frequency.
132
LIST SWEEP ONLY
When a sweep type other than the list sweep was specified, you attempted to execute a
command to specify all segments (entire sweep range) as the search range setting
(SEARNG FULL) or a command to specify a certain segment as the search target
(SEGMNUM <value>) and, as a result, the command was ignored. Before setting the
search range for the list sweep, specify the list sweep as the sweep type (command: SWPT
LIST).
133
CAN'T CHANGE ON LIST SWEEP
When the sweep type was set to the list sweep, you attempted to execute a command to
498