Outpdmkr - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

OUTPDMKR?

Syntax
OUTPDMKR?
Description
For the active trace (set with the "TRAC" command on page 457), reads out the
measurement parameter value and sweep parameter value of the Δ marker position. If the Δ
marker mode is off, executing this command causes an error and the readouts are invalid.
(Query only)
Query response
{numeric 1},{numeric 2},{numeric 3}<newline><^END>
Readouts are as follows:
{numeric 1}:
{numeric 2}:
{numeric 3}:
Corresponding
No front panel key is available to execute this function.
key
Chapter 16
Readout of the measurement parameter value at the Δ marker position.
If the measurement parameter is a scalar value (for other than
COMPLEX Z-Y), the measurement parameter value is read out. If the
measurement parameter is a vector value (for COMPLEX Z-Y), the
readout is as follows:
If the polar chart format is selected, the amplitude is read out.
If the complex plane format is selected, resistance (for trace A) or
conductance (for trace B) is read out.
Subsidiary readout of the measurement parameter value at the
Δ marker position. If the measurement parameter is a scalar value (for
other than COMPLEX Z-Y), 0 is always read out. If the measurement
parameter is a vector value (for COMPLEX Z-Y), the readout is as
follows:
If the polar chart format is selected, the phase is read out.
If the complex plane format is selected, reactance (for trace A) or
susceptance (for trace B) is reads out.
Sweep parameter value at the Δ marker position.
GPIB Command Reference
OUTPDMKR?
377