LIMUSTAR

Syntax
LIMUSTAR <numeric>[DEG|RAD|OHM|SIE|H|F|PCT]
LIMUSTAR?
Description
In the creation/edit of the limit line table, sets the upper limit of the start point of each
segment (specified with the "LIMSTAR" command on page 334). Use this command,
together with the "LIMLSTAR" command on page 328, when specifying the limits using
the upper limit and the lower limit.
If segment creation/edit has not been started, executing this command automatically starts
the edit of the edit target segment and then sets the upper limit of the start point.
If the creation/edit of the limit line table has not been started, executing this command
automatically starts table creation/edit, starts the edit of the edit target segment, and then
sets the upper limit of the start point.
Parameters
Description
Range
Initial value
Unit
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[System]
key
Chapter 16
Upper limit of the segment start point
-1E9 to 1E9
When creating segment 1: 0
When creating an additional segment: Set value of the previous
segment
Varies depending on the measurement parameter (refer to the
explanation of unit in "SCAL" on page 423).
-
-
LIMIT TEST
EDIT LIMIT LINE
GPIB Command Reference
<numeric>
-
EDIT - UPPER LIMIT
LIMUSTAR
337