Rplpps; Rplrhei; Rsco - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

RPLPPS?

RPLPPS?
Syntax
RPLPPS?
Description
Within the waveform analysis range specified with the "ANARANG" command on page
267, reads out the difference between the maximum peak measurement parameter value
and the minimum negative peak measurement parameter value and the sweep parameter
value of the peaks (if there are several maximum peaks or minimum negative peaks, the
value of the leftmost peak within the analysis range). If the value cannot be obtained
because no peak or negative peak is detected, 0 is read out.
The peak and negative peak used at the execution of this command satisfies the condition
set with the "THRR" command on page 452. It is not related to the peak definition in the
marker search function. (Query only)
Query response
{numeric 1},{numeric 2},{numeric 3}<newline><^END>
Readouts are as follows:
{numeric 1}:
{numeric 2}:
{numeric 3}:
Corresponding
No front panel key is available to execute this function.
key

RPLRHEI?

Syntax
RPLRHEI?
Description
Within the waveform analysis range specified with the "ANARANG" command on page
267, calculates the right ripple (measurement parameter value difference between the peak
and its right adjacent negative peak), and reads out the maximum value. If the value cannot
be obtained because no peak or negative peak is detected, 0 is read out.
The peak and negative peak used at the execution of this command satisfies the condition
set with the "THRR" command on page 452. It is not related to the peak definition in the
marker search function. (Query only)
Query response
{numeric}<newline><^END>
Corresponding
No front panel key is available to execute this function.
key

RSCO

Syntax
RSCO
Description
Returns the tint (set with the "TINT" command on page 454), brightness (set with the
"CBRI" command on page 276), and chroma (set with the "COLOR" command on page
281) of the display color of the item selected with the "COLO" command on page 280 to
the factory states. (No query)
Corresponding
[Display]
key
-
414
Difference between the maximum peak measurement parameter value
and the minimum negative peak measurement parameter value.
Sweep parameter value of the maximum peak.
Sweep parameter value of the minimum negative peak.
-
-