Dcalshor{R|L - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

DCALSHOR{R|L}

Syntax
DCALSHORR <numeric>[OHM]
DCALSHORL <numeric>[H]
DCALSHOR{R|L}?
Description
For user calibration data measurement ("CAL{A|B|C}" command on page 274), sets the
resistance value ("DCALSHORR" command) and the inductance value ("DCALSHORL"
command) of the SHORT standard of the calibration kit used to measure SHORT
calibration data.
Parameters
Description
Range
Initial value
Unit
Resolution
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[CAL]
key
Chapter 16
Resistance value or inductance value
-1E6 to 1E6
0
For resistance value: Ω (ohm)
For inductance value: H (henry)
1E-12
-
USER CAL
-
DEFINE VAUE
GPIB Command Reference
<numeric>
- {
SHORT RESIST (R)
|
SHORT INDUCT (L)
DCALSHOR{R|L}
}
289

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents