GPIB Command Reference

PAVER

PAVER
Syntax
PAVER {ON|OFF|1|0}
PAVER?
Description
Enables/disables the point averaging function.
Parameters
Query response
{1|0}<newline><^END>
Corresponding
[Bw/Avg]
key

PAVERFACT

Syntax
PAVERFACT <numeric>
PAVERFACT?
Description
Sets the point averaging count when using the point averaging function. To set the point
averaging count of each segment when creating the list sweep table, also use this
command.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Bw/Avg]
key
396
ON or 1
Enables the point averaging function.
OFF or 0
Disables the point averaging function.
(initial value)
-
POINT AVG on OFF
Description
Point averaging count
Range
1 to 256
Initial value
4 (Note that, if you use this command to create a segment, the
initial value for segment 1 is 1, and the initial value of an additional
segment is the set value of the previous segment.)
Resolution
1
-