GPIB Command Reference

OUTPLIMM?

OUTPLIMM?
Syntax
OUTPLIMM?
Description
For the active trace (set with the "TRAC" command on page 457), reads out the limit test
result of measurement point of the marker position. If the marker function is off, executing
this command causes an error and invalid values are read out. (Query only)
Query response
{numeric 1},{numeric 2},{numeric 3},{numeric 4}<newline><^END>
The readouts are as follows:
Corresponding
No front panel key is available to execute this function.
key

OUTPMAX?

Syntax
OUTPMAX?
Description
Reads out the maximum measurement parameter value and the sweep parameter value at
the position, within the waveform analysis range set with the "ANARANG" command on
page 267. If there are several measurement points of the maximum measurement parameter
value, reads out the value of the leftmost measurement point within the analysis range.
(Query only)
Query response
{numeric 1},{numeric 2}<newline><^END>
Readouts are as follows:
{numeric 1}:
{numeric 2}:
Corresponding
No front panel key is available to execute this function.
key
384
{numeric 1}: Sweep parameter value
{numeric 2}: Limit test result
0: FAIL
1: PASS
-1: Test was off.
{numeric 3}: Upper border value of the limit
{numeric 4}: Lower border value of the limit
Maximum measurement parameter value.
Sweep parameter value at the position of the maximum measurement
parameter value.
Chapter 16