Trgs; Uskey - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

TRGS

Syntax
TRGS {INT|EXT|BUS|MAN}
TRGS?
Description
Selects a trigger source.
Parameters
INT (initial value)
EXT
BUS
MAN
Query response
{INT|EXT|BUS|MAN}<newline><^END>
Corresponding
[Trigger]
key

USKEY

Syntax
USKEY
Description
Displays user keys (ON KEY LABELS) of Instrument BAISC. (No query)
Corresponding
[System]
key
Chapter 16
Specifies the internal trigger.
Specifies the external trigger inputted from the EXT TRIGGER
terminal on the rear panel.
Specifies the GPIB/LAN trigger (trigger by executing the "*TRG"
command on page 261).
Specifies the manual trigger (trigger by the following key sequence
on the front panel:
-
SOURCE [ ]
-{
FREE RUN
-
IBASIC
-
ON KEY LABELS
GPIB Command Reference
Description
[Trigger]
-
SOURCE [ ]
-
|
EXTERNAL
|
HP-IB/LAN
|
TRGS
MANUAL
).
MANUAL
}