Clel; Cles - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

CLEL

Syntax
CLEL
Description
In the creation/edit of the list sweep table, deletes all segments. Executing this command
during the execution of segment creation/edit causes an error. If list sweep table
creation/edit has not been started, executing this command automatically starts table
creation/edit and then deletes all segments. (No query)
NOTE
Executing the "EDITDONE" command on page 310 brings the created/edited list sweep
table to take effect.
Corresponding
[Sweep]
key

CLES

Syntax
CLES
Description
Clears the error queue, Status Byte Register, Operation Status Register, Standard Event
Status Register, and Instrument Event Status Register. This command has the same
function as the "*CLS" command on page 258. (No query)
Corresponding
No front panel key is available to execute this function.
key