GPIB Command Reference

COMST{A|B|C}

COMST{A|B|C}
Syntax
COMST{A|B|C} {ON|OFF|1|0}
COMST{A|B|C}?
Description
Toggles on and off the OPEN/SHORT/LOAD compensation of the fixture compensation
function. The "COMSTA" command is used to set the OPEN compensation; the
"COMSTB" command, the SHORT compensation; the "COMSTC" command, the LOAD
compensation. If no fixture compensation coefficient has been stored (executed with the
"COM{A|B|C}" command on page 281), executing this command to turn on the
compensation causes an error and the command is ignored.
Parameters
Query response
{1|0}<newline><^END>
Corresponding
[Cal]
key

CONT

Syntax
CONT
CONT?
Description
Sets the sweep mode to the auto continuous sweep (CONT). In this mode, sweeps are
repeated automatically and continuously.
Query response
{1|0}<newline><^END>
Corresponding
[Trigger]
key

COPA

Syntax
COPA
Description
Aborts printout. (No query)
Corresponding
[Copy]
key
282
ON or 1
OFF or 0 (Initial value) Turns off the compensation.
-
FIXTURE COMPEN
1 (Initial value) The auto continuous sweep is selected as the sweep mode.
0
The auto continuous sweep is not selected as the sweep mode.
-
CONTINUOUS
-
ABORT
Turns on the compensation.
- {
OPEN on OFF
|
SHORT on OFF
Description
Description
|
LOAD on OFF
}