GPIB Command Reference

SPAN

SPAN
Syntax
SPAN <numeric>[HZ|MHZ|V|A]
SPAN?
Description
Sets the span value of the sweep range. To set the span value of each segment when
creating the list sweep table, also use this command.
Parameters
Note that, if you use this command to create a segment, the initial value is 0 regardless of
the sweep parameter.
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Span]
key
442
Description
When the sweep
parameter is
frequency
When the sweep
parameter is
OSC level (voltage)
When the sweep
parameter is
OSC level (current)
When the sweep
parameter is
dc bias level (voltage)
When the sweep
parameter is
dc bias level (current)
Span value
Range
0 to 109.99996E6 (for linear sweep)
20 to 109.99996E6 (for log sweep)
Initial value 109.99996E6
Unit
Hz
Resolution 1E-3
Range
0 to 0.995
Initial value 0.995
Unit
V (volt)
Resolution 1E-3
Range
0 to 19.8E-3
Initial value 19.8E-3
Unit
A (ampere)
Resolution 10E-6
Range
0 to 80
Initial value 0
Unit
V (volt)
Resolution 1E-3
Range
0 to 200E-3
Initial value 0
Unit
A (ampere)
Resolution 10E-6
<numeric>
Chapter 16