SPLD

Syntax
SPLD {ON|OFF|1|0}
SPLD?
Description
Enables/disables the vertically separate display for traces A and B.
If the separate display of traces A and B is enabled, the upper half of the LCD screen is
used as trace A measurement result display screen, and the lower half, as trace B
measurement result display screen. If the separate display of traces A and B is disabled, the
measurement results of traces A and B are superimposed on the same screen.
If the screen is set so that the upper half of the LCD screen is used as the measurement
result display screen and the lower half as the HP Instrument BASIC screen (specified to
HIHB with the "DISA" command on page 298), regardless of the setting to enable/disable
the separate display, both traces are superimposed on the same screen.
Parameters
ON or 1
OFF or 0
(initial value)
Query response
{1|0}<newline><^END>
Corresponding
[Display]
key
Chapter 16
Enables the separate display of traces A and B.
Disables the separate display of traces A and B.
-
SPLIT on OFF
GPIB Command Reference
Description
SPLD
443