E4Tp; Ecal{P|A|B|C; Ecaldon - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

E4TP

Syntax
E4TP {OFF|M1|M2|APC7|PROBE}
E4TP?
Description
Selects the adapter in the adapter setting.
Parameters
OFF (initial value)
M1
M2
APC7
PROBE
Query response
{OFF|M1|M2|APC7|PROBE}<newline><^END>
Corresponding
[Cal]
key

ECAL{P|A|B|C}

Syntax
ECAL{P|A|B|C}
Description
Measures the PHASE/OPEN/SHORT/LOAD data to calculate the setup data of the
selected adapter. The "ECALP" command measures the PHASE data; the "ECALA"
command, the OPEN data; the "ECALB" command, the SHORT data; the "ECALC"
command, LOAD data. These data must be measured for each adapter (selected with the
"E4TP" command on page 309). (No query)
Corresponding
[Cal]
key

ECALDON

Syntax
ECALDON
Description
Finishes the measurement of the data for the adapter setup and stores the measured data
into the non-volatile memory. The setup data is stored as the data for the adapter selection
specified at the measurement (data is stored for each adapter selection). (No query)
Corresponding
[Cal]
key
Chapter 16
Specifies no adapter (NONE).
Specifies the Agilent 16048G (4TP 1M).
Specifies the Agilent 16048H (4TP 2M).
Specifies the Agilent 42942A (7mm 42942A).
Specifies the Agilent 42941A (PROBE 42941A).
-
- {
|
ADAPTER [ ]
NONE
4TP 1M
-
ADAPTER [ ]
-
SETUP
- {
PHASE COMP [ ]
-
-
-
ADAPTER [ ]
SETUP
done
GPIB Command Reference
Description
|
|
|
4TP 2M
7mm 42942A
PROBE 42941A
|
OPEN [ ]
|
SHORT [ ]
E4TP
}
|
LOAD [ ]
}
309