GPIB Command Reference
PEAKCENT
PEAKCENT
Syntax
PEAKCENT
Description
For the active trace (set with the "TRAC" command on page 457), searches for a peak
using the marker and changes the sweep center value setting to the marker sweep
parameter value (in other words, the sweep parameter value of the detected peak). If the
marker function is off, executing this command causes an error and the command is
ignored. (No query)
Corresponding
[Marker→]
key
PHAU
Syntax
PHAU {DEG|RAD}
PHAU?
Description
For the active trace (set with the "TRAC" command on page 457), sets the display unit in
the phase display format. The scale parameters in the selected unit are automatically set so
that traces do not exceed the display range on the screen.
Parameters
Query response
{DEG|RAD}<newline><^END>
Corresponding
[Format]
key
398
PEAK → CENTER
-
-
more 1/2
DEG (initial value) Specifies degree (°) as the display unit.
RAD
Specifies radian as the display unit.
-
PHASE UNIT [ ]
Description
Chapter 16