When the delta marker was not displayed, you attempted to execute a function that needed
the display of the delta marker and, as a result, the command was ignored. For example,
when the delta marker was not displayed, a command corresponding to the MKRD?SPAN
key (MKRDSPAN) was sent. Before executing a function using the delta marker, first
display the delta marker (command: DMKR ON).
100
NO FIXED DELTA MARKER
When the fixed Δ marker (FIXED DMKR) was not displayed, you attempted to execute a
command of a function that needed the display of the fixed Δ marker and, as a result, the
command was ignored. For example, when the fixed Δ marker was not displayed, a
command corresponding to the FIXED DMKR VALUE, FIXED DMKR AUX VALUEN
key (DMKRVAL <value>, DMRRAUV <value>) was sent. Before executing a function using
the fixed Δ marker, first display the fixed Δ marker (command: DMKR FIX).
101
SEARCH WIDTH OFF
When the band width function was off, you attempted to execute a command to search the
cutoff point on the trace (WIDSIN or WIDSOUT) and, as a result, the command was
ignored. Before executing the cutoff point search of the band width function, turn on the
band width function (command: WIDT ON).
102
SEARCH RANGE TYPE IS NOT SEGMENT
Though, in the list sweep, all segments (entire sweep range) were specified as the search
range (command: SEARNG FULL), you attempted to execute a command to specify a
certain segment as a search target (SEGMNUM <value>) and, as a result, the command
was ignored. Before executing the search function by specifying a certain list sweep
segment, set the search range to a single segment (command: SEARNG SEGMENT).
105
MEASUREMENT PARAMETER NOT FOR ANALYSIS
When a measurement parameter other than |Z|-q was set, you attempted to execute a
waveform analysis command (for example, OUTPCERR?) and, as a result, the command
was ignored. Before executing a waveform analysis command, set the measurement
parameter to |Z|-θ (command: MEAS IMPH).
110
SAVE ERROR
When saving a file, anomalies in the storage media were detected. For example, if you
attempt to save a file on a floppy disk, it may be damaged. Check to see if there is no
problem in the storage media.
111
RECALL ERROR: INSTR STATE PRESET
Because an error occurred during reading out a file, the 4294A is being preset. This error
occurs, for example, when data in a file to be read out is destroyed or when an extension for
a file name does not match with the contents of the file.
112
INVALID FILE NAME
When the recall command (RECD <string>) or the resave command (RESAVD <string>)
was executed, a string that indicated a file name was not followed by a necessary extension
and, as a result, the command was ignored. Add a required extension, and then perform
recall or resave. This message occurs, for example, when you attempt to execute the recall
or resave command without adding an extension (".STA", ".DAT", ".TXT") to a file name.
Also, it does when you attempt to execute resave of a graphics file without adding an
extension (".TIF").
113
NO STATE/DATA FILES ON DISK
Appendix D
Error messages
Error number: 100
497