SDEL

Syntax
SDEL [<numeric>]
Description
In the creation/edit of the list sweep table, deletes the specified segment (if no segment is
specified, the segment specified as the edit target) from the table.
The segment number of each segment after the deleted segment decreases by 1. As a result,
the segment, whose number was changed to the same number as the deleted segment, is set
to the edit target. Note that, if the last segment is deleted, the segment immediately before
the deleted segment is set to the edit target. You can set any segment to the edit target using
the "SEGM" command on page 435.
During segment edit (after the execution of the "SADD" command on page 415 or "SEDI"
command on page 434 and before the execution of the "SDON" command on page 426),
executing this command causes an error and the command is ignored.
If the creation/edit of the list sweep table has not been started, executing this command
automatically starts table creation/edit and then deletes the segment. (No query)
NOTE
Executing the "EDITDONE" command on page 310 brings the created/edited list sweep
table to take effect.
Parameters
Description
Range
Resolution
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Corresponding
[Sweep]
key
Chapter 16
Segment number you want to delete
1 to the number of segments in the list sweep table
1
-
-
EDIT LIST
DELETE
GPIB Command Reference
<numeric>
SDEL
425