Agilent Technologies 4294A Programming Manual page 223

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

3100
CASE ELSE
3110
GOTO Quit
3120
END SELECT
3130 Prog_end: END
3140
!
3150
! Correction Data Measurement Function
3160
!
3170
DEF FNUser_Corr(@Agt4294a,Type$)
3180
DIM Inp_char$[9],Err_mes$[50]
3190
INTEGER Err_no
3200
!
3210
OUTPUT @Agt4294a;"ESNB 256"
3220
OUTPUT @Agt4294a;"*SRE 4"
3230
OUTPUT @Agt4294a;"*CLS"
3240
!
3250
ON INTR 8 GOTO Meas_end
3260
ENABLE INTR 8;2
3270
SELECT Type$
3280
CASE "Adapter_Phase"
3290
OUTPUT @Agt4294a;"ECALP"
3300
CASE "Compen_Open"
3310
OUTPUT @Agt4294a;"COMA"
3320
CASE "Compen_Short"
3330
OUTPUT @Agt4294a;"COMB"
3340
CASE "Compen_Load"
3350
OUTPUT @Agt4294a;"COMC"
3360
END SELECT
3370
PRINT "Now measuring..."
3380 Meas_wait:
GOTO Meas_wait
3390 Meas_end:
!
3400
OUTPUT @Agt4294a;"OUTPERRO?"
3410
ENTER @Agt4294a;Err_no,Err_mes$
3420
IF Err_no=0 THEN
3430
PRINT Standard$&" Data Measurement Complete"
3440
RETURN 0
3450
ELSE
3460
PRINT "Error: "&Err_mes$
3470
PRINT "Program Interruption"
3480
OUTPUT @Agt4294a;"ECALQUI"
3490
RETURN -1
3500
END IF
3510
FNEND
13
Application Sample Programs
Measuring Magnetic Materials
223