TINT

TINT
Syntax
TINT <numeric>
TINT?
Description
Sets the hue of the display color of the item selected with the "COLO" command on page
280 command.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Display]
key
-

TITL

Syntax
TITL <string>
TITL?
Description
For the active trace (set with the "TRAC" command on page 457), sets a string displayed in
the title area on the display screen.
Parameters
If the specified file name has characters that exceed the maximum length, only the string of
up to the maximum length is valid and the remaining characters are ignored.
Query response
{string}<newline><^END>
Corresponding
[Display]
key
454
Description
Value indicating the hue (0 or 100: red, 33: green, 66: blue)
Range
0 to 100
Initial value
Varies depending on the display item selected with the "COLO"
command on page 280.
Resolution
1
-
-
more 1/2
MODIFY COLOR
TINT
Description
String displayed in the title area
Range
65 characters or less
Initial value
Blank ("")