Meas al Options Menu
he analyzer uses array of calibration data to improve the measurements made by the
hardware. his menu allow you to disable the calibration, and view the calibration data arrays.
: Determines whether cal arrays are used in processing data.
Meas Cal ON off
will switch OFF both user-performed cals and the default factory cal. his
Meas Cal OFF
includes transmission, re ection, and frequency response cal arrays (both broadband and
narrowband). his function provides a way to examine the uncorrected performance of the
hardware in the analyzer.
Detect Cal ON off
external broadband detectors.
View Array Master
displayed. hese are the cal arrays from the currently active calibration (default, full band, or
user-de ned).
View Array Interpol
hese are the cal arrays from the currently active calibration (default, full band, or
user-de ned). If the frequency range of the analyzer has not been changed since the last
calibration was performed, then this is the same as the master.

Analog Bus Menu

he analyzer has an analog bus that can be used to troubleshoot hardware problems on some
of the boards in the analyzer. his is done by sampling analog bus nodes, which provide
measurements at selected points on the boards. Nominal values for each node are provided in
able 5-5.
: Allows you to select an analog bus node to be measured. he currently
elect Node
selected node description and number are displayed. A node may be selected by using the
keypad to enter the number, or by using the front-panel knob or the
through the list.
: riggers a data measurement of the selected analog bus node.
ample
: Allows you to switch OFF the power cal for both the internal and
: Selects the master (uninterpolated) cal arrays from the master to be
: Selects the interpolated cal arrays from the master to be displayed.
*
+
/
keys to scroll
5
5
Serv ce Related Menus
5-9