OUTPVACP?

Syntax
OUTPVACP? <numeric 1>
Description
When the oscillator level monitor function is set to on (set to ON with "OMON" command
on page 365), reads out the oscillator voltage level monitor value of the specified
measurement point. (Query only)
Parameters
Description
Range
Resolution
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric 2}<newline><^END>
Description
If the oscillator level monitor function is set to off, the readout is invalid.
Corresponding
No front panel key is available to execute this function.
key

PARS

Syntax
PARS {ON|OFF|1|0}
PARS?
Description
For the marker search function for the active trace (set with the "TRAC" command on page
457), determines whether to perform partial search.
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Corresponding
[Search]
key
Chapter 16
Number of the measurement point you want to read out (the
number assigned in order from the left edge assuming that the
number of the leftmost measurement point is 1)
1 to the number of points
1
Voltage monitor value of the oscillator level
Enables the partial search.
Disables the partial search.
-
-
SEARCH RANGE MENU
GPIB Command Reference
<numeric 1>
{numeric 2}
Description
PART SRCH on OFF
OUTPVACP?
395