GPIB Command Reference

STOP

STOP
Syntax
STOP <numeric>[HZ|MHZ|V|A]
STOP?
Description
Sets the sweep range stop value.
Parameters
Note that, if you use this command to create a segment, the initial value when creating
segment 1 is the current sweep range start value and the initial value when creating an
additional segment is the sweep range stop value of the previous segment.
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Stop]
key
446
Description
When the sweep
parameter is
frequency
When the sweep
parameter is
OSC level (voltage)
When the sweep
parameter is
OSC level (current)
When the sweep
parameter is
dc bias level (voltage)
When the sweep
parameter is
dc bias level (current)
Stop value
Range
40 to 110E6 (for linear sweep)
60 to 110E6 (for log sweep)
Initial value 110E6
Unit
Hz
Resolution 1E-3
Range
5E-3 to 1
Initial value 1
Unit
V (volt)
Resolution 1E-3
Range
200E-6 to 20E-3
Initial value 20E-3
Unit
A (ampere)
Resolution 10E-6
Range
-40 to 40
Initial value 0
Unit
V (volt)
Resolution 1E-3
Range
-100E-3 to 100E-3
Initial value 0
Unit
A (ampere)
Resolution 10E-6
<numeric>
Chapter 16