Outperro; Outpfaip; Outpiac - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

OUTPERRO?

OUTPERRO?
Syntax
OUTPERRO?
Description
Reads out the oldest error among errors stored in the error queue of the 4294A. The size of
the error queue is 10. (Query only)
Query response
{numeric},{string}<newline><^END>
If no error is stored in the error queue, the read-out error number is 0 and the error message
is "No error."
Corresponding
No front panel key is available to execute this function.
key

OUTPFAIP?

Syntax
OUTPFAIP?
Description
For the active trace (set with the "TRAC" command on page 457), reads out the number of
points whose limit test result is FAIL. (Query only)
Query response
{numeric}<newline><^END>
Corresponding
No front panel key is available to execute this function.
key

OUTPIAC?

Syntax
OUTPIAC?
Description
When the oscillator level monitor function is set to on (set to ON with "OMON" command
on page 365), reads out the oscillator current level monitor values of all measurement
points. (Query only)
Query response
{numeric 1},{numeric 2},..,{numeric NOP}<newline><^END>
Where, NOP is the number of points.
If the oscillator level monitor function is set to off, the readouts are invalid.
Corresponding
No front panel key is available to execute this function.
key
380
{numeric}
Description Error number
{string}
Error message (string with double quotation marks ("))
Chapter 16

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents