Swed; Swet - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

SWED

SWED
Syntax
SWED {UP|DOWN}
SWED?
Description
Selects the sweep direction.
Parameters
Query response
{UP|DOWN}<newline><^END>
Corresponding
[Sweep]
key

SWET

Syntax
SWET <numeric>[S]
SWET?
Description
Sets the sweep time.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Sweep]
key
448
UP (initial value)
Specifies the sweep parameter increasing direction (from left to
right on the screen).
DOWN
Specifies the sweep parameter decreasing direction (from right to
left on the screen).
-
DIRECTION [ ]
Description
Sweep time
Range
0 to 10000 (Note that, the lower limit varies depending on the
measurement condition.)
Initial value
Varies depending on the measurement condition.
Unit
s (second)
Resolution
0.001
-
-
TIME
SWEEP TIME
Description
<numeric>
Chapter 16

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents