POSL

Syntax
POSL
POSL?
Description
Sets the input/output signals of the 24-bit I/O port (data and test PASS/FAIL output) to
positive logic. To set to negative logic, use the "NEGL" command on page 362.
Query response
{1|0}<newline><^END>
1
0
Corresponding
No front panel key is available to execute this function.
key

POWE

Syntax
POWE <numeric>[V|A]
POWE?
Description
Sets the oscillator (OSC) power level. To set the oscillator power level of each segment
when creating the list sweep table, also use this command. To select voltage or current to
set the level, use the "POWMOD" command on page 404.
Parameters
Description
Range
Initial value
Unit
Resolution
Note that, if you use this command to create a segment, the initial value when creating
segment 1 is the current oscillator power level value and the initial value when creating an
additional segment is the oscillator power level value of the previous segment.
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Source]
key
Chapter 16
Positive logic is specified.
Positive logic is not specified. In other words, negative logic (preset
state) is specified.
Oscillator power level
For voltage setting: 5E-3 to 1
For current setting: 200E-6 to 20E-3
For voltage setting: 0.5
For current setting: 200E-6
For voltage setting: V (volt)
For current setting: A (ampere)
For voltage setting: 1E-3
For current setting: 10E-6
-
LEVEL
GPIB Command Reference