GPIB Command Reference
DCI
DCI
Syntax
DCI <numeric>[A]
DCI?
Description
Sets the dc bias output level when the dc bias output mode is the current mode or
constant-current mode. To set the dc bias output level of each segment when creating the
list sweep table, also use this command.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Source]
key
290
Description
Output current value of dc bias
Range
-0.1 to 0.1
Initial value
0 (Note that the initial value is the current set value of the dc bias
output current, when creating segment 1; the set value of the
previous segment, when creating an additional segment.)
Unit
A (ampere)
Resolution
20E-6
-
-
BIAS MENU
<numeric>
CURRENT LEVEL
Chapter 16