RPLHEI?

Syntax
RPLHEI?
Description
Within the waveform analysis range specified with the "ANARANG" command on page
267, calculates ripples (measurement parameter value difference between the peak and its
right and left adjacent negative peaks), and reads out the maximum value. If the value
cannot be obtained because no peak or negative peak is detected, 0 is read out.
The peak and negative peak used at the execution of this command satisfies the condition
set with the "THRR" command on page 452. It is not related to the peak definition in the
marker search function. (Query only)
Query response
{numeric}<newline><^END>
Corresponding
No front panel key is available to execute this function.
key

RPLLHEI?

Syntax
RPLLHEI?
Description
Within the waveform analysis range specified with the "ANARANG" command on page
267, calculates the left-side ripple (measurement parameter value difference between the
peak and its left adjacent negative peak), and reads out the maximum value. If the value
cannot be obtained because no peak or negative peak is detected, 0 is read out.
The peak and negative peak used at the execution of this command satisfies the condition
set with the "THRR" command on page 452. It is not related to the peak definition in the
marker search function. (Query only)
Query response
{numeric}<newline><^END>
Corresponding
No front panel key is available to execute this function.
key

RPLPP?

Syntax
RPLPP?
Description
Within the waveform analysis range specified with the "ANARANG" command on page
267, reads out the difference between the maximum peak measurement parameter value
and the minimum negative peak measurement parameter value. If the value cannot be
obtained because no peak or negative peak is detected, 0 is read out.
The peak searched for at the execution of this command satisfies the condition set with the
"THRR" command on page 452. It is not related to the peak definition in the marker search
function. (Query only)
Query response
{numeric}<newline><^END>
Corresponding
No front panel key is available to execute this function.
key
Chapter 16
GPIB Command Reference
RPLHEI?