Pkdlty; Pkpol - Agilent Technologies 4294A Programming Manual

Precision impedance analyzer
Hide thumbs Also See for 4294A:
Table of Contents

Advertisement

GPIB Command Reference

PKDLTY

PKDLTY
Syntax
PKDLTY <numeric>[OHM|DEG|RAD|SIE|H|F|PCT]
PKDLTY?
For the active trace (set with the "TRAC" command on page 457), sets the ΔY value (refer
Description
to Figure 7-1 on page 97) parameter to define the peak searched for in the marker search
function.
Parameters
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Search]
key

PKPOL

Syntax
PKPOL {POS|NEG}
PKPOL?
Description
For the active trace (set with the "TRAC" command on page 457), sets the polarity of the
peak searched for in the marker search function.
Parameters
Query response
{POS|NEG}<newline><^END>
Corresponding
[Search]
key
400
ΔY value
Description
Range
0 to 100E6
Initial value
1
Unit
Varies depending on the measurement parameter (refer to the
explanation of unit in "SCAL" on page 423).
-
PEAK
-
PEAK DEF MENU
POS (initial value)
Specifies positive peaks.
NEG
Specifies negative peaks.
-
-
PEAK
PEAK DEF MENU
<numeric>
PEAK DEF: ΔY
-
Description
-
PEAK PLRTY [ ]
Chapter 16