Advertisement

Preparing For Accurate Measurement
Adapter Setting
200
!
210
Result=FNAdap_setup(@Hp4294a,Scode,"Phase")
220
IF Result<>0 THEN Prog_end
230
!
240
Result=FNAdap_setup(@Hp4294a,Scode,"Open")
250
IF Result<>0 THEN Prog_end
260
!
270
Result=FNAdap_setup(@Hp4294a,Scode,"Short")
280
IF Result<>0 THEN Prog_end
290
!
300
Result=FNAdap_setup(@Hp4294a,Scode,"Load")
310
IF Result<>0 THEN Prog_end
320
!
330
OUTPUT @Hp4294a;"ECALDON"
340
OUTPUT @Hp4294a;"*OPC?"
350
ENTER @Hp4294a;Buff$
360
PRINT "All Data Measurement Complete"
370
!
380 Prog_end: END
390
!
400
! Adapter Setup Data Measurement Function
410
!
420
DEF FNAdap_setup(@Hp4294a,INTEGER Scode,Standard$)
430
DIM Inp_char$[9],Err_mes$[50]
440
INTEGER Err_no
450
OUTPUT @Hp4294a;"*CLS"
460
PRINT "Set "&Standard$&"-Connection"
470
INPUT "OK? [Y/N]",Inp_char$
480
IF UPC$(Inp_char$)="Y" THEN
490
ON INTR Scode GOTO Meas_end
500
ENABLE INTR Scode;2
510
SELECT Standard$
520
CASE "Phase"
530
OUTPUT @Hp4294a;"ECALP"
540
CASE "Open"
550
OUTPUT @Hp4294a;"ECALA"
560
CASE "Short"
570
OUTPUT @Hp4294a;"ECALB"
580
CASE "Load"
590
OUTPUT @Hp4294a;"ECALC"
600
END SELECT
610
PRINT "Now measuring..."
620 Meas_wait:
GOTO Meas_wait
630 Meas_end:
!
640
OUTPUT @Hp4294a;"OUTPERRO?"
650
ENTER @Hp4294a;Err_no,Err_mes$
660
IF Err_no=0 THEN
670
PRINT Standard$&" Data Measurement Complete"
680
RETURN 0
690
ELSE
700
PRINT "Error: "&Err_mes$
710
PRINT "Program Interruption"
720
OUTPUT @Hp4294a;"ECALQUI"
730
RETURN -1
740
END IF
750
ELSE
760
PRINT "Program Interruption"
770
OUTPUT @Hp4294a;"ECALQUI"
56