MANS

Syntax
MANS {ON|OFF|1|0}
MANS?
Description
Turns on/off the manual sweep function.
Parameters
ON or 1
OFF or 0 (initial value)
Query response
{1|0}<newline><^END>
Corresponding
[Sweep]
key

MAXDCV

Syntax
MAXDCV <numeric>[V]
MAXDCV?
Description
Sets the voltage limit maximum value when the dc bias output mode is the constant-voltage
mode or constant-current mode (CVOLT or CCURR has been specified with the
"DCMOD" command on page 291).
Parameters
Description
Range
Initial value
Unit
Resolution
If the specified parameter is out of the allowable setting range, the minimum value (if the
lower limit of the range is not reached) or the maximum value (if the upper limit of the
range is exceeded) is set.
Query response
{numeric}<newline><^END>
Corresponding
[Source]
key
Chapter 16
Turns on the manual sweep function.
Turns off the manual sweep function.
-
MANUAL SWP on OFF
Voltage limit maximum value of the dc bias output
-40 to 40
40
V (volt)
1E-3
-
-
BIAS MENU
MAX LIMIT VORTAGE
GPIB Command Reference
Description
<numeric>
MANS
347